Maintenance
— 38
SPECTRO GENESIS FES 27 — 17.07.2019
6.5
Checking the aerosol formation of the nebulizer
In order to proceed with the inspection, follow the steps below:
1.
Remove the nebulizer, see page 35, but without disconnecting the argon
connection.
2.
Call up the “Solid State Generator” menu item via the instrument control
parameter dialog.
3.
Adjust the nebulizer gas flow determined for the nebulizer.
4.
Using a spray bottle, force
distilled water through the
nebulizer connection for sample
introduction.
The spray coming out must
be evenly V-shaped. If this
is not the case, the
nebulizer may be clogged
and must be cleaned, see
page 35.
Fig. 12
5.
Reinstall the components.
Checking of aerosol generation is now complete.