Safety
— 14
SPECTRO GENESIS FES 27 — 17.07.2019
Warning of hand injury!
The instrument is equipped with a peristaltic pump with rotating parts.
Ensure to keep out of harm's way while the pump is running. All object must
be kept clear of the pump.
When conveying hazardous substances, the appropriate safety precautions
must be taken for each respective liquid and application.
Warning! Risk of suffocation!
The instrument is operated with argon. Argon is heavier than air. In case of
leakage, the argon can accumulate in a closed room, especially on the floor or
lower-lying areas.
Ensure to provide adequate ventilation. If necessary, install oxygen sensors.
Warning! Optical radiation!
The plasma emits UV radiation. A protective pane is used to shield from the
radiation.
Operate the instrument only with the protective pane in place and the door
closed.
Information
A high-frequency emitter is used to operate the ICP, dependent on the type of
model. The electromagnetic emission is below the legally specified limit values.
Please point out to your employees that an electromagnetic emitter is used by
attaching the cardiac pacemaker symbol!
Information
In order to ensure the availability and operational safety of the equipment, we
recommend having the system checked for correct operation and performance at
least once per year by specialized technicians. To find contact addresses of
service organizations authorized by SPECTRO go to www.spectro.com.