Technical data
SPECTRO GENESIS FES 27 — 17.07.2019
— 15
3
Technical data
Description
Value
Instrument height
870 mm; 34.3"
Instrument width
1,165 mm; 45.9"
Instrument depth
748 mm; 29.5"
Instrument weight
approx. 115 kg; ~ 254 lbs.
Required space
1070 x 680 mm; 42.2 x 26.8"
Power supply
230 V AC ± 5 %, 50/60 Hz
Phase + N PE, the instrument is
equipped with a CEE 32A plug.
Instrument power consumption
3.5 kVA
Instrument required line protection
30 – 32 A (inertial)
Sound pressure level
58 dBA
Stability
(according to DIN EN 61010-1)
compliant if change of position <10°
Storage conditions
2–40°C; 36–104°F
Operating range (permissible ambient
temperature)
5 – 35 °C; 41 – 95°F
Recommended temperature range for
reaching the specified performance
18–25°C; 64–77°F, maximum temperature
gradient ∆T = 3°C/h (5.4°F/h)
Atmospheric humidity
< 80% rel. (non-condensing)
Air quality
Free of corrosive components and high dust
concentrations
Operation
Max. 3000 m above mean sea level
Summary of Contents for FES 27