Maintenance
SPECTRO GENESIS FES 27 — 17.07.2019
— 37
7.
Separate the nebulizer from the spray chamber by slightly twisting it.
8.
Disconnect the waste water connection from the drainage of the spray
chamber and remove the cap from the spray chamber by slightly turning it.
9.
Clean the spray chamber.
First, use a suitable diluted acid solution (e.g., hydrochloric acid or nitric
acid 1:10).
Hazardous substance!
Observe the corresponding information on hazardous materials.
After this, use a laboratory cleaning agent for removing traces of fatty
substances (streaks) if required.
Now flush the spray chamber thoroughly with distilled water.
10. Reinstall the components.
11. Optimize the nebulizer gas flow, see page 41.
Cleaning of the spray chamber is now complete.