6 – Installation and set-up
OxyDetect Oxygen Deficiency Gas Detector
40
05311001B / Revision 4
Parameter
Values/options
Range
The selected option determines the mA output range
associated with a measurement:
Range 1
The output is set to use range 1
Range 2
The output is set to use range 2
To avoid accidental changes between range 1 and range 2 they
should be set at the same levels. Range 1 should be used and the
option to switch between ranges password protected.
Pour éviter les modifications accidentelles entre la plage et la
plage 1 2 ils doivent être réglés au même niveau. Range 1 doit
être utilisé et la possibilité de basculer entre les gammes mot de
passe protégé.
Range 1 low level
The range 1 lowest sample measurement
Range 1 high level
The range 1 highest sample measurement (span)
Range 2 low level
The range 2 lowest sample measurement
Range 2 high level
The range 2 highest sample measurement (span)
During calibration
The selected option determines how the mA output will
operate during calibration:
Freeze
As soon as the calibration screen is displayed,
the mA output will 'freeze' at its last output
value. The output will only be updated to reflect
subsequent measurements when calibration
screen has been exited.
Follow
The mA output value will reflect the
measurement value, even during calibration.
Jam condition
The selected option determines how the mA output will
operate during a fault condition:
High
The output value will be held at 21.5 mA
(suitable for trace measurements)
Low
The output value will be held at 0 mA
(suitable for purity measurements)