Fault symptom
Recommended actions
The measurement alarms are
activating more often than
expected.
Check that the instrument is not being knocked,
moved, or subjected to high levels of vibration
during sample measurements.
Check that the alarm modes, alarm levels and
hysteresis levels have been correctly set: refer to
Section 6.5.
The milliamp output is at 0 or
21.5 mA.
If you have configured the mA output to jam high
or jam low, check whether a fault condition exists.
Otherwise, contact Servomex or your local
Servomex agent for assistance.
The milliamp output is not as
expected.
Ensure that the electrical cabling connected to the
instrument is not open circuit.
Check that the mA output is calibrated correctly
(see Section 6.4.4).
Check that you have selected the correct Range
(see Section 6.4).
A relay signal output is not as
expected.
Check that the signal cable is correctly connected
to the analyser: refer to Section 6.2.5.