6 – Installation and set-up
OxyDetect Oxygen Deficiency Gas Detector
38
05311001B / Revision 4
6.3.6 Setting the clock
The time and date will remain set for approximately 1 week if the power
supply to the instrument is switched off. After this period, it will be necessary
to reset the date and time in order to clear the
date/time invalid
fault condition.
Use the following procedure to set the date and time:
1.
Select
Settings
Clock.
2.
Edit the displayed time as described in Section 5.10.
Figure 18 - The clock 9time) screen
3.
Edit the displayed date as described in Section 5.10.
4. The date format can be set to your regional preference ('day/month/year' or
'month/day/year'): refer to Section 6.3.7.
Figure 19 - The clock (date) screen
6.3.7 Changing regional settings
You can configure the following instrument regional settings so that the information
shown on the various screens is better suited to your local conventions:
Setting
Options available
Language
Various languages are supported
Date format
Day/Month/Year * or Month/Day/Year
Decimal format
Use of "." * (full stop) or "," (comma) as the decimal point
* Default option.