6 – Installation and set-up
OxyDetect Oxygen Deficiency Gas Detector
30
05311001B / Revision 4
1. Connect the wires in your cable to the screw terminals on TB1 see Figure 12.
Terminal
Use
TB1-1
+24V dc
TB1-2
0V dc
Table 1 – Electrical supply connection
TB1 dc power supply connection cable specifications
Cable type
Unscreened
Number of cores required
2
Rating (temperature)
+5 °C to +45 °C (+41 °F to +113 °F)
Flammability rating
VW-1
Approvals
Relevant to local requirements
Cable external diameter
Within the range specified for the
selected cable gland
Conductors
0.2 to 2.5mm
2
/ 26 to 12 AWG
Strip length 7mm
Terminal screws
0.4 to 0.5 Nm tightening torque
Screwdriver blade
0.6 x 3.5, PH 0, PZ 0
Table 2 - TB1 dc power supply connection cable specifications
6.2.4 Analogue output connections
The terminals of all equipment connected to the OxyDetect by
the user shall be separated from mains voltages by at least
reinforced insulation.
Les bornes de tous les équipements connectés à l'OxyDetect
par l'utilisateur doivent être séparés des tensions d'alimentation
par une isolation au moins renforcée.