05311001B / Revision 4
39
To change the regional settings:
1.
Select
Settings
Regional.
2.
This screen shows the first regional option (Language).
Figure 20 - The regional settings (language) option screen
6.4
Configuring and using the mA outputs (option)
6.4.1 Overview
The instrument is supplied factory configured with a 4-20mA output scaled 0-25% for
the gas measurement on both range 1 & 2.
The mA output provides a constantly updated output, in which the current represents
the value of the gas sample measurement.
The instrument allows you to specify two separate range configurations per
measurement for the mA outputs: range 1 and range 2. The current range is shown
on the measurement screen (see Figure 6):
is shown if range 1 is selected.
is shown if range 2 is selected.
The mA output can be selected as:
0 to 20 mA, where 0 mA represents the lowest sample measurement and 20 mA
represents the highest sample measurement in the range you have specified.
4 to 20 mA, where 4 mA represents the lowest sample measurement and 20 mA
represents the highest sample measurement in the range you have specified.
In addition to the above, you can specify how the mA output will operate during
calibration, fault conditions and under-range conditions. Details of the output
parameters for the mA outputs are given in Section 6.4.2. Set up, configure, check,
calibrate and use the mA outputs as described in Section 6.4.4.
6.4.2 Introduction to the mA output parameters
The mA output parameters that you must set up are as follows: