80
issued after the test is started until the end of the test or interrupted by other commands.
FETCh
To get the measurement result.
--
Syntax:
Command message:
FETCh: AUTO <ON/OFF> or <1/0>
Query message:
FETCh:AUTO?
--
Data<ON/OFF> or <1/0>
Data format: character
Data range: 0 (OFF), 1 (ON)
--
Example:
Set auto returned test result to ON.
Command message: FETCh:AUTO ON
Query message: FETCh:AUTO?
Return value: ON
FETCh?
Output the test result.
Command Syntax: FETCh? After the tester receives this command, the tester will automatically issue
the test results for all steps until the end of the test.
Return Format: test steps: test mode, actual test voltage (kV) ,actual test current (A), test result;
Example: STEP 1:AC,1.000,1.000e-3,PASS; STEP 2:DC,1.500,0.100e-3,PASS;
Test Steps: 1, test mode: AC, actual test voltage 1kV, actual test current 1mA, test result:PASS;
Test Steps: 2, test mode: DC, actual test voltage1.5kV, actual test current 0.1mA, test result:PASS;
Note: The tester will automatically return each test result by default (per step test result).
5.2.8 Other Commands
*IDN
Inquiry about the model and version information of the instrument.
Return message: <manufacturer>,<model>,<firmware><NL^END>
Where,
<manufacturer> Gets the manufacturer name (Sourcetronic)
<model> Gets the machine model (such as ST9110/9110A)
<firmware> Gets the version number of firmware (such as Version1.0.5)
For example: *IDN?
FUNC:START
Start Test
*STOP
Stop Test