64
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the rise time for IR test in STEP1 to 1s.
Command message: FUNC:SOUR:STEP 1:IR:RTIM 1
Query message: FUNC:SOUR:STEP 1:IR:RTIM?
Return value: 1.0
FUNC:SOURce:STEP:IR:FTIM
To set /inquiry about the FALL time for IR test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:IR:FTIM<time>
Query message: FUNC:SOUR:STEP <sn>:IR:FTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the fall time for IR test in STEP1 to 1s.
Command message: FUNC:SOUR:STEP 1:IR:FTIM 1
Query message: 1.0
FUNC:SOURce:STEP:IR:RANG
To set /inquiry about the range for IR test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:IR:RANG<range>
Query message: FUNC:SOUR:STEP <sn>:IR:RANG?
--
Data<range>:
Data format: integer
Data range: 0~6 (0 is AUTO, 1 is 10mA, 2 is 3mA, 3 is 300uA, 4 is 30uA, 5 is 3uA, 6 is
300nA)
--
Example:
Set the range for IR test in STEP1 to10mA.
Command message: FUNC:SOUR:STEP 1:IR:RANG 1
Query message: FUNC:SOUR:STEP 1:IR:RANG?
Return value: 1
5.2.3.6 Commands for PA Setup Function
FUNC:SOURce:STEP:PA:MESSAge
To set /inquiry about the PA message.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:PA:MESSAge <message >
Query message: FUNC:SOUR:STEP <sn>:PA:MESSAge?
--
Data<message>: