70
--
Syntax:
Command message: FUNC:BREAK:RTIM<time>
Query message: FUNC:BREAK:RTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the rise time for breakdown test to1s.
Command message: FUNC:BREAK:RTIM 1
Query message: FUNC:BREAK:RTIM?
Return value: 1.0
FUNC: BREAKdown:WTIM
To set /inquiry about the WAIT TIME for voltage breakdown
test.
--
Syntax:
Command message:FUNC:BREAK:WTIM<time>
Query message: FUNC:BREAK:WTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the waiting time for breakdown test to1s.
Command message: FUNC:BREAK:WTIM 1
Query message: FUNC:BREAK:WTIM?
Return value: 1.0
FUNC:BREAK:ARC
To set /inquiry about the ARC high current for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:ARC<current >
Query message: FUNC:BREAK:ARC?
--
Data<current>:
Data format: float
Data range: 0, 1.0~10.0mA (DC), 1.0 ~ 20mA (AC) (0 is OFF)
Data accuracy: 0.1mA
Data unit: mA
--
Example:
Set the arc upper current for breakdown test to 5mA.
Command message: FUNC:BREAK:ARC 5
Query message: FUNC:BREAK:ARC?
Return value: 5.0