68
Data range: 0 ~ 1 (0: AC; 1: DC)
--
Example:
Set the mode for breakdown test: 1 (DC)
Command message: FUNC:BREAK:MODE 1
Query message: FUNC:BREAK: MODE?
Return value: 1
FUNC:BREAKdown:VOLTStart
To set /inquiry about the VOLTAGE START for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:VOLTS<voltage >
Query message: FUNC: :BREAK:VOLTS?
--
Data<voltage>:
Data format: float
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the starting voltage for breakdown test to 1000V.
Command message: FUNC:BREAK:VOLTS 1000
Query message: FUNC:BREAK:VOLTS?
Return value: 1000
FUNC:BREAKdown:VOLTEnd
To set /inquiry about the VOLTAGE END for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:VOLTE<voltage>
Query message: FUNC: :BREAK:VOLTE?
--
Data<voltage>:
Data format: float
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the ending voltage for breakdown test to 5000V.
Command message: FUNC:BREAK:VOLTE 5000
Query message: FUNC:BREAK:VOLTE?
Return value: 5000
FUNC: BREAKdown:UPPC
To set /inquiry about the HIGH CURRENT for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:UPPC <current>
Query message: FUNC:BREAK:UPPC?