57
Query message:
FUNC:SOUR:STEP <sn>:AC:TTIM?
--
Data<time>:
Data format: float
Data range: 0, 0.3~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the TEST time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:TTIM 1
Query message: FUNC:SOUR:STEP 1:AC:TTIM?
Return message: 1.0
FUNC:SOURce:STEP:AC:RTIM
To set /inquiry about the RISE time for ACW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:RTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:RTIM?
--
Data<time>:
Data format: float
Data range: 0~999.9s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the RISE time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:RTIM 1
Query message: FUNC:SOUR:STEP 1:AC:RTIM?
Return message: 1.0
FUNC:SOURce:STEP:AC:FTIM
To set /inquiry about the FALL time for ACW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:FTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:FTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the FALL time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:FTIM 1