67
Data format: integer
Data range: 0 ~ 1 (0: normal step; 1: voltage step)
--
Example:
Set the step mode for breakdown test to 1 (voltage step).
Command message: FUNC:BREAK:STEPMODE 1
Query message: FUNC:BREAK: STEPMODE?
Return value: 1
FUNC:BREAKdown:STEP
To set /inquiry about the STEP for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:STEP<step>
Query message: FUNC: :BREAK: STEP?
--
Data<step>:
Data format: integer
Data range: 2~999
Data accuracy: 1
--
Example:
Set the step for breakdown test: 55
Command message: FUNC: :BREAK: STEP 55
Query message: FUNC: :BREAK: STEP?
Return value: 55
FUNC:BREAKdown:STEPVOLT
To set /inquiry about the STEP VOLTAGE for voltage
breakdown test .
--
Syntax:
Command message: FUNC:BREAK:STEOVOLT<step voltage >
Query message: FUNC: :BREAK: STEOVOLT?
--
Data<step voltage >:
Data format: float
Data range: 5~500
Data accuracy: 1
Data unit: V
--
Example:
Set the step voltage for breakdown test to 50V.
Command message: FUNC:BREAK: STEOVOLT 50
Query message: FUNC:BREAK: STEOVOLT?
Return value: 50
FUNC:BREAKdown:MODE
To set /inquiry about the MODE for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:MODE<test mode>
Query message: FUNC: :BREAK: MODE?
--
Data<mode>:
Data format: integer