29
MeasRang
: Select the appropriate current range for
IR
, calculate the current value according to the
test voltage and the expected insulation impedance of the DUT, ie I = U/R, and then select the
appropriate current range accordingly. Relationship between the current range and resistance
measurement range is shown in table
4-6.
Current Range
Resistance Measurement Range
Set Voltage
50V~499V
Set Voltage
500V~5000V
10mA(3~10mA)
0.1MΩ~1MΩ
0.1MΩ~4.5MΩ
3mA(0.3~3mA)
0.5MΩ~4.5MΩ
3.0MΩ~15.0MΩ
300uA(30~300µA)
3.0MΩ~15.0MΩ
10.0MΩ~45MΩ
30uA(3~30µA)
10.0MΩ~45MΩ
35.0MΩ~450MΩ
3uA(0.3~3µA)
45MΩ~0.45GΩ
0.40GΩ~4.5GΩ
300nA(20~300nA)
0.40GΩ~4.9GΩ
4.0GΩ~50.0GΩ
Table 4-6
4.3.4 Pause Mode
Figure 4-15 Pause Mode Setup
Message: message prompt string, you can enter English letters, Arabic numerals, and [.] and [-] in
the status bar up to a maximum of 20 characters, after which ST will appear in the 21th position.
Press keys from 2 to 9 repeatedly to cycle through the letters depicted on them. You can also use the
softkeys to add characters. Use [BAS] to erase the last letter, finish with [ENTER], cancel with
[ESC] or a cursor key.
PausTime: the action mode of setting the pause mode
a)
When set to 0, to continue testing, the pause mode would be ended as either the [START] key
on the front panel is pressed, or the external [START] signal is retriggered.
b) When set to 0.3~999.0s, if the set time finishes, the pause mode ends.