69
--
Data<current>:
Data format: float
Data range: 0.0001~25mA (DC); 0.001~100mA (AC)
Data accuracy: 0.0001mA (DC); 0.001mA (AC)
Data unit: mA
--
Example:
Set the upper current for breakdown test to10mA.
Command message: FUNC:BREAK:UPPC 10
Query message: FUNC:BREAK:UPPC?
Return value: 10.000
FUNC: BREAKdown:LOWC
To set /inquiry about the LOW CURRENT for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:LOWC<current>
Query message: FUNC:BREAK:LOWC?
--
Data<current>:
Data format: float
Data range: 0~uppder current (0 is OFF)
Data accuracy: 0.0001mA (DC); 0.001mA (AC)
Data unit: mA
--
Example:
Set the lower current for breakdown test to1mA.
Command message: FUNC:BREAK:LOWC 1
Query message: FUNC:BREAK:LOWC?
Return value: 1.000
FUNC: BREAKdown:TTIM
To set /inquiry about the TEST TIME for voltage breakdown
test.
--
Syntax:
Command message: FUNC:BREAK:TTIM<time>
Query message: FUNC:BREAK:TTIM?
--
Data<time>:
Data format: float
Data range: 0, 0.3~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the test time for breakdown test to1s.
Command message: FUNC:BREAK:TTIM 1
Query message: FUNC:BREAK:TTIM?
Return value: 1.0
FUNC: BREAKdown:RTIM
To set /inquiry about the RISE TIME for voltage breakdown test.