59
Data range: 1~50
Data accuracy: 1
--
Data<voltage>:
Data format: float
Data range: 50~6000
Data accuracy: 1
Data unit: V
--
Example:
Set the voltage for DCW test in STEP1 to 1000V
Command message: FUNC:SOUR:STEP 1:DC:VOLT 1000
Query message: FUNC:SOUR:STEP 1:DC:VOLT?
Return value: 1000
FUNC:SOURce:STEP:DC:UPPC
To set /inquiry about the UPPER current for DCW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:DC:UPPC <current>
Query message: FUNC:SOUR:STEP <sn>:DC:UPPC?
--
Data<current>:
Data format: float
Data range:
0.0001~25.000mA (voltage>=1500V)
0.0001~20.000mA (voltage<1500V)
Data accuracy: 0.0001mA
Data unit: mA
--
Example:
Set the upper current for DCW test in STEP1 to 1mA
Command message: FUNC:SOUR:STEP 1:DC:UPPC 1
Query message: FUNC:SOUR:STEP 1:DC:UPPC?
Return value: 1.000
FUNC:SOURce:STEP:DC:LOWC
To set /inquiry about the LOWER current for DCW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:DC:LOWC<current>
Queery message:FUNC:SOUR:STEP <sn>:DC:LOWC?
--
Data<current>:
Data format: float
Data range: 0~upper current (0 is OFF)
Data accuracy: 0.0001mA
Data unit: mA
--
Example:
Set the lower current for DCW test in STEP1 to 1mA
Command message: FUNC:SOUR:STEP 1:DC:LOWC 1
Query message: FUNC:SOUR:STEP 1:DC:LOWC?
Return value: 1.000