6.Command Specification
REQUEST SENSE
SONY AIT-3 drive SDX-700 series Ver.2.1
6-94
44 00
INTERNAL TARGET FAILURE - Any unexpected internal error
CONDITIONS INDICATING A NON-RECOVERABLE HARDWARE FAILURE.
REFER TO THE FIELD REPLACEABLE UNIT CODE FIELD (SENSE BYTE 14) TO
DETERMINE WHICH PART OF THE DRIVE IS FAILING. HOST SHOULD ISSUE
SEND DIAGNOSTICS WITH THE SELF-TEST BIT SET TO THOROUGHLY TEST
THE DRIVE. THIS SENSE INFORMATION COULD ALSO BE RETURNED AFTER A
SEND DIAGNOSTICS COMMAND ENDS WITH A CHECK CONDITION STATUS.
HOST SHOULD ISSUE RECEIVE DIAGNOSTICS RESULTS FOR MORE DETAILS.
POSSIBLE CAUSES: BUFFER OVERRUN/ UNDER-RUN, FORMAT VIOLATION,
MECHANICAL STATUS TIME-OUT, INTERNAL DATA PARITY ERROR, BYTE
COUNT MISMATCH, INCORRECT MICRO PROCESSOR OPERATION,
UNEXPECTED SCSI PROTOCOL CHIP INTERRUPT.
83 06
MIC READ ERROR
05h
ILLEGAL REQUEST
. Indicates that there was an illegal parameter in the Command
Descriptor Block or in the additional parameters supplied as data for some commands.
If the target detects an invalid parameter in the command descriptor block, then it shall
terminate the command without altering the medium. If the target detects an invalid
parameter in the additional parameters supplied as data, then the target may have
already altered the medium. This sense key may also indicate that an invalid
IDENTIFY message was received.
1A 00
PARAMETER LIST LENGTH ERROR - Indicates that the Parameter Length error in
the Command Descriptor Block. Either the value does not fall on a Header, Page or
Block descriptor boundary or the Parameter List Length is unexpectedly large.
20 00
INVALID COMMAND OPERATION CODE - Invalid or Unsupported command.
24 00
INVALID FIELD IN CDB – Bits are set in one or more fields are not supported or are
Reserved. Refer to the Field Pointer (Bytes 15-17) to identify the illegal bit or field.
25 00
LOGICAL UNIT NOT SUPPORTED - The SDX-700C has an embedded SCSI
controller with only one Logical Unit (LUN). The error indicates that the LUNTRN field
in the Identify message was not ZERO.
26 00
INVALID FIELD IN PARAMETER LIST - Test Number,
Configuration Number Header, or Field in the Parameter List was invalid or selected
Diagnostic test is not Host assessable.
26 01
PARAMETER NOT SUPPORTED - Unsupported page requested. Refer to the Field
Pointer (Bytes 15-17) to identify the illegal bit or field.
26 02
PARAMETER VALUE INVALID
2C 00
COMMAND SEQUENCE ERROR - Write Buffer command when tape is inserted and
not at BOT.
39 00
SAVING PARAMETERS NOT SUPPORTED
3D 00
INVALID BITS IN IDENTIFY MESSAGE - Identify message was not 80h or C0h.
55 04
INSUFFICIENT REGISTRATION RESOURCES(attempt to register more than 1 key)
83 02
No MIC Detected
83 04
MIC out of Memory
83 05
No Data Body Detected in the MIC
83 0C
MIC format type is undefined
83 0D
MIC capacity field indicates illegal value.
83 80
Partition Count exceeded
Summary of Contents for SDX-700C
Page 12: ...1 Introduction SONY AIT 3 drive SDX 700 series Ver 2 1 1 4 This page intentionally left blank ...
Page 28: ...3 Installation SONY AIT 3 drive SDX 700 series Ver 2 1 3 6 This page intentionally left blank ...
Page 178: ...10 Appendix C 10 2 SONY AIT 3 drive SDX 700 series Ver 2 1 This page intentionally left blank ...
Page 180: ...11 Appendix D 11 2 SONY AIT 3 drive SDX 700 series Ver 2 1 This page intentionally left blank ...
Page 186: ...12 Appendix E SONY AIT 3 drive SDX 700 series Ver 2 1 12 6 This page intentionally left blank ...