14
Chapter 3:
Troubleshooting
Failure (Assert) Codes
The system displays an “assert screen” for hardware and software issues related to main PCBA failures. Main PCBA
failures typically result in “assert codes” that are output to the display. If an assert screen appears, note the assert
code and contact SonoSite Technical Support to clarify the failure.
shows an assert screen. The assert
code is the bracketed number on the line labeled “C:”.
Figure 3.1 Assert Screen
Verifying a System Assert Code
System asserts are caused by hardware and/or software faults. Hardware asserts typically require main PCBA
replacement. Software asserts can be reset and the system may recover. A simple method to identify the cause of
the assert is identified here:
Assert code
Assert Cause
1
Record the assert code.
2
Press and release the
Power
button to power the system down.
3
Press the
Power
button again to power on the system.
• If the system powers on normally, it has recovered from the fault (software assert) and
you may use the system.
• If the assert condition remains, corrective action must be taken; usually replacement of
the main PCBA is required. Contact SonoSite Technical Support for assistance and to
obtain repair parts.
If the
Power
button is not functional, all sources of power must be removed to allow the
system to power down. I.e., disconnect AC power and remove the battery.
Summary of Contents for M-Turbo
Page 1: ...M Turbo Ultrasound System Service Manual TM...
Page 6: ...2 Chapter 1 Introduction...
Page 20: ...16 Chapter 3 Troubleshooting...
Page 36: ...32 Chapter 4 Replacement Procedures...
Page 44: ...40 Chapter 5 Performance Testing...
Page 52: ...48 Appendix B Service Event Report Service Event Report Form...
Page 56: ...52 Index...
Page 57: ......
Page 58: ...P08144 01 P08144 01...