36
Chapter 5:
Performance Testing
Additional Performance Tests
Color Doppler (Color)
Color Power Doppler (CPD)
Table 5.2: Imaging Performance
Imaging
Performance
C11x
C60x
ICTx
HFL38
L25x
L38x
P21x
Exam type
Nerve
OB
OB
Small
Parts
Sup
Breast
ABD
Optimization
Gen
Gen
Gen
Res
Res
Res
Pen
2D Penetration
6.8cm
14.0 cm
6.5 cm
4.5 cm
4.3 cm
5.7 cm
21.0 cm
Test Color
1
Connect any transducer.
2
Press the
Color
key. “Color” should be annotated in the top left corner of the display.
3
A Region of Interest (ROI) box is displayed on top of the grayscale image. Use the touchpad
to move the CPD ROI. Verify that the ROI moves to the new position on the display.
4
Adjust the
Depth
control for minimum depth in the image.
5
Adjust the
Gain
control so that color speckles just appear inside the ROI box.
6
Gently tap the face of the transducer and observe that the ROI box fills with color
information.
7
Press the
Freeze
key and then save the image. Press the
Freeze
key again to return to live
imaging.
Test CPD
1
Connect any transducer.
2
Press the
Color
key. A Region of Interest (ROI) box is displayed on top of the grayscale image.
3
Press the
Color
softkey to switch to CPD. “CPD” should be annotated in the top left corner
of the display.
4
Adjust the
Depth
control for minimum depth in the image.
5
Adjust the
Gain
control so that color speckles just appear inside the ROI box.
6
Gently tap the face of the transducer and observe that the ROI box fills with color
information.
Summary of Contents for M-Turbo
Page 1: ...M Turbo Ultrasound System Service Manual TM...
Page 6: ...2 Chapter 1 Introduction...
Page 20: ...16 Chapter 3 Troubleshooting...
Page 36: ...32 Chapter 4 Replacement Procedures...
Page 44: ...40 Chapter 5 Performance Testing...
Page 52: ...48 Appendix B Service Event Report Service Event Report Form...
Page 56: ...52 Index...
Page 57: ......
Page 58: ...P08144 01 P08144 01...