Chapter 4:
Replacement Procedures
19
5
Disconnect the two connectors from the display to the Main PCBA per
.
Figure 4.4 Display Connectors
6
The replacement Display Assembly does not include the Display Rear Enclosure. Remove
the Display Rear Enclosure by removing the two screw caps, two screws, and then sliding
the rear enclosure up and away from the display as show in
.
Figure 4.5 Remove Display Back Enclosure
Connectors (2)
Remove Screw Caps and
Screws (2)
Display rear enclosure slides
up