background image

 

-57- 

 

PS##-OMM0003-A 

 

 

Z/ISE80

-C01L

 

 

 

Z/ISE80

-A2

 

 

Summary of Contents for ISE80(H)

Page 1: ...No PS OMM0003 A PRODUCT NAME Digital Pressure Switch MODEL Series Product Number ZSE80 F ISE80 H...

Page 2: ...tion 25 F1 Setting of OUT1 26 F2 Setting of OUT2 29 F3 Setting of response time 29 F4 Setting of analog output auto shift input 30 F5 Setting of display resolution 32 F7 Setting of fine adjustment of...

Page 3: ...equipment failure when configuring the equipment 2 Only personnel with appropriate training should operate machinery and equipment The product specified here may become unsafe if handled incorrectly T...

Page 4: ...acement product or necessary parts will be provided This limited warranty applies only to our product independently and not to any other damage incurred due to the failure of the product 3 Prior to us...

Page 5: ...double interlocking system for example a mechanical system Check the product regularly for proper operation Otherwise malfunction can result causing an accident The following instructions must be fol...

Page 6: ...ten the lifetime of the Pressure switch Design the product to prevent reverse current when the circuit is opened or the product is forced to operate for operational check Reverse current can cause mal...

Page 7: ...lar never lift a Pressure switch equipped with fitting and piping by holding the lead wires Otherwise damage to the internal parts can result causing malfunction or to be off the connector Avoid repea...

Page 8: ...expose the product to direct sunlight If using in a location directly exposed to sunlight shade the product from the sunlight Otherwise failure or malfunction can result Keep within the specified flui...

Page 9: ...f water or drainage getting in narrow the diameter of the piping to the Pressure switch or make an orifice in the middle of the piping Although this Pressure switch uses stainless diaphragm pressure s...

Page 10: ...Description Piping direction Part number ZS 24 A Rear surface piping ZS 24 D Bracket Bottom piping ZS 35 A Rear surface piping ZS 35 C Panel mount adapter Bottom piping ZS 35 B Rear surface piping ZS...

Page 11: ...d error code Four display modes can be selected display always in red or green only or changing from green to red linked to output button UP Selects the mode or increases ON OFF set value Press this b...

Page 12: ...frequency C Chattering preventing function A function to delay the response time of switch output in order to prevent chattering D Digit Min setting unit Shows how precisely the pressure can be indica...

Page 13: ...n L Load impedance See Max load impedance Manual setup Manual pressure setup without using auto preset This term is used to discriminate from the pressure setup using auto preset Max applied voltage T...

Page 14: ...and current consumption is reduced Pressure sensing part A pressure detecting part of a pressure detecting element Pressure setting The setting of pressure to determine the point at which the Pressure...

Page 15: ...unction Function to change the unit in which the value of pressure is indicated Only a product with this function can change the unit A product with unit changing function cannot be purchased if it is...

Page 16: ...When the Pressure switch is to be mounted in a place where water and dust splashes occur insert a tube O D 4 I D 2 5 into the air relieving port of the Pressure switch Mounting by bracket Fix the brac...

Page 17: ...16 PS OMM0003 A Mounting with panel mount adapter Rear piping Bottom piping...

Page 18: ...l mount adapter Panel mount adapter A and B included Model ZS 35 B Bottom piping ZS 35 C Rear piping Panel mount adapter Front protective cover Model ZS 35 E Bottom piping ZS 35 F Rear piping Panel th...

Page 19: ...htening torque of 13 6 Nm or less Tube attachment When this Pressure switch is used in a place where water and dust splash on insert a tube in the air relieving port and bring piping of the opposite s...

Page 20: ...ilter and ferrite core between the switch mode power supply and the product or by using a series power supply instead of a switch mode power supply Internal circuit and wiring example N NPN open colle...

Page 21: ...nalog output 4 to 20 mA Max load impedance Power supply voltage 12 V 300 Power supply voltage 24 V 600 Min load impedance 50 V T Auto shift input mode With auto shift switch output PNP open collector...

Page 22: ...ds a set point the Pressure switch will be turned ON When the pressure falls below the set point by the amount hysteresis or more the Pressure switch will be turned OFF The default setting of the Pres...

Page 23: ...Press the button to finish the setting For models with 2 outputs P_2 or n_2 will be displayed Set as above If the button is pressed for 2 seconds or longer the setting is fixed and measurement mode r...

Page 24: ...ings are provided If the setting is acceptable keep it for use To change setting enter function selection mode F 0 Unit conversion function See page 25 Symbol Pressure range Default setting ISE80 H MP...

Page 25: ...e See page 29 2 5 ms F 4 Setting of analog output auto shift input See page 30 Analog output F 5 Setting of display resolution See page 32 1000 split F 7 Setting of fine adjustment of display value Se...

Page 26: ...function selection mode to display F 0 Press the button Move on to unit conversion function Press the button to set Return to function selection mode Setting of F 0 unit conversion function completed...

Page 27: ...of output modes on page 28 Operation Press the or button in function selection mode to display F 1 Press the button Move on to setting output mode Press the button to set Move on to setting of revers...

Page 28: ...F 1 operation of OUT1 completed 1 Selected parameter become effective after pressing the button 2 After having setting effective by the button it is possible to move to measurement mode by pressing th...

Page 29: ...tput is changed is out of the set pressure range due to selection between normal and reversed output hysteresis is automatically compensated The above figure shows the operation at OUT1 For the operat...

Page 30: ...of OUT1 page 26 to 28 F 3 Setting of response time Chattering prevention function Select the response time of the switch output Output chattering is prevented by setting the response time Operation Pr...

Page 31: ...ue to zero referring to the pressure when signal is input Operation Press the or button in function selection mode to display F 4 Press the button Move on to setting of function Press the button to se...

Page 32: ...it or lower limit which is closer of the set pressure range automatically The correction is performed when the auto shift input is applied at the pressure beyond the set pressure range If the auto shi...

Page 33: ...f display resolution Press the button to set Return to function selection mode Setting of F 5 display resolution completed Not selectable depending on selected pressure unit The display resolution sel...

Page 34: ...button Move on to setting of display resolution Press the button to set Press the button Move on to setting of adjustment initialize Press the button to set Return to function selection mode Setting o...

Page 35: ...ton in function selection mode to display F 8 Press the button Move on to setting of auto preset Press the button to set Return to function selection mode Setting of F 8 auto preset completed Press th...

Page 36: ...nd AP2 will appear 4 Selection of auto preset OUT2 Press the button to set P_1 and H_1 n_1 and H_1 for reverse mode and display AP2 If OUT2 does not need to be set press the and buttons simultaneously...

Page 37: ...button to set Return to function selection mode Setting of F 9 power saving mode completed In power saving mode a key operation returns to normal display and if there is no key operation for 30 secon...

Page 38: ...e Press the button to set Return to function selection mode Setting of F10 security code completed If the security code is used it is necessary to input the security code to release the key lock The s...

Page 39: ...eturn to function selection mode After the change to oFF disabled press the button to set Return to function selection mode Press the button for 2 seconds or longer 1 Setting of functions Every time b...

Page 40: ...tting of response time All models 12 Setting of grey wire function 13 Setting of auto shift function 14 Setting of effective output Only output type of R S T and V 15 Setting of display resolution All...

Page 41: ...ss the button Move on to reset to the default setting oFF unused selected Press the button to set Return to function selection mode The setting is reset to the default setting and the mode returns to...

Page 42: ...r A displayed value can be adjusted to zero when the pressure to be measured is within 10 of the span between atmospheric pressure and upper limit of rated pressure range from the default value of pre...

Page 43: ...he or button to select locking of the key UnL 3 When the button is pressed the input of security code is asked For how to input the security code refer to How to input and change the security code on...

Page 44: ...ge the security code The changed security code is indicated 3 After checking it is as desired press the button for 1 second or longer The measurement mode is returned At this time if the or button is...

Page 45: ...setting UnL or LoC is indicated Keep pressing and buttons simultaneously for 5 seconds or longer Then keep pressing and buttons simultaneously for 5 seconds or longer There is no change of the indica...

Page 46: ...ressure switch does not operate normally Refer to reference No 1 The switch output is on The indication light is on The indication light works improperly The switch output is off The switch output gen...

Page 47: ...ference No 11 Refer to reference No 14 Refer to reference No 13 Refer to reference No 15 Refer to reference No 16 Refer to reference No 17 Refer to reference No 10 The indication accuracy does not sat...

Page 48: ...the selected model output type Lead wire breakage Check if there is bending stress applied to any parts of the lead wire Bending radius and tensile force applied to the lead wire Correct the wiring co...

Page 49: ...itch Er0 4 6 and 7 1 Check if there is noise interference such as static electricity Check if there is a noise source 2 Check if the power supply voltage is in the range of 12 to 24 VDC 10 1 Remove th...

Page 50: ...Reset the setting of function Incorrect wiring Check the power supply wiring Check if the brown and blue wires are connected to DC and DC respectively and if the output line is about to come off cont...

Page 51: ...Japan due to a new measurement law It is fixed to the SI unit kPa MPa 13 The unit cannot be changed Product failure Replace the product Key lock mode Check if the key lock mode is turned on Turn off...

Page 52: ...nces Perform zero clear operation again after restoring the applied pressure to an atmospheric pressure condition Pressure has exceeded the upper limit of the set pressure range Pressurizing Error Pre...

Page 53: ...0 mA load current Response time 2 5 ms With anti chatter function 20 100 500 1000 2000 ms selectable Short circuit protection Provided Repeatability 0 2 F S 1 digit Hysteresis mode Hysteresis Window c...

Page 54: ...vinyl cabtyre cable 3 cores N P 3 5 2 m 4 cores A B Sectional area of conductor 0 15 mm2 AWG26 5 cores R T S V Outside diameter of insulator 0 95 mm Standard CE UL CSA RoHS Piping specifications Model...

Page 55: ...54 PS OMM0003 A Dimensions Z ISE80 02 Z ISE80 02L...

Page 56: ...55 PS OMM0003 A Z ISE80 N02 Z ISE80 N02L...

Page 57: ...56 PS OMM0003 A Z ISE80 F02 Z ISE80 C01...

Page 58: ...57 PS OMM0003 A Z ISE80 C01L Z ISE80 A2...

Page 59: ...58 PS OMM0003 A Z ISE80 A2L Z ISE80 B2...

Page 60: ...59 PS OMM0003 A Z ISE80 B2L...

Page 61: ...PS OMM0003 A Revision history A Revision Note Specifications are subject to change without prior notice and any obligation on the part of the manufacturer 2008 2010 SMC Corporation All Rights Reserved...

Reviews: