-23-
PS##-OMM0003-A
Setting of Function
■
Function selection mode
In measurement mode, press the
button for 2 seconds or longer to display [F 0].
Select the display of function setting to be changed, [F
].
Press the
button for 2 seconds or longer in function selection mode to return to measurement mode.
∗
: Some functions are not available depending on part number. All functions are displayed with [F
] and followed with function
description. If a function is not available for specified type, the function is displayed as [- - -].
■
Default setting
At the time of shipment, the following settings are provided.
If the setting is acceptable, keep it for use.
To change setting, enter function selection mode.
•
[F 0] Unit conversion function
See page 25
Symbol
Pressure range
Default setting
ISE80(H) MPa
Nil or M
ZSE80(F) kPa
ISE80(H)
P
ZSE80(F)
psi
•
[F 1] Parameter setting of OUT1
See page 26
Item
Explanation
Default setting
Output mode
Selects hysteresis mode or window comparator mode.
Hysteresis mode
Reversed output Selects reversed output.
Normal output
Pressure setting
Sets ON or OFF point of the switch output.
ISE80: 0.500 MPa
ZSE80: -50.5 kPa
ZSE80F: 50.0 kPa
ISE80H: 1.000 MPa
Hysteresis
Chattering can be prevented by setting hysteresis.
ISE80: 0.500 MPa
ZSE80: -50.5 kPa
ZSE80F: 50.0 kPa
ISE80H: 1.000 MPa
Display colour
Selects the display colour.
ON: Green
OFF: Red
Summary of Contents for ISE80(H)
Page 17: ...16 PS OMM0003 A Mounting with panel mount adapter Rear piping Bottom piping...
Page 55: ...54 PS OMM0003 A Dimensions Z ISE80 02 Z ISE80 02L...
Page 56: ...55 PS OMM0003 A Z ISE80 N02 Z ISE80 N02L...
Page 57: ...56 PS OMM0003 A Z ISE80 F02 Z ISE80 C01...
Page 58: ...57 PS OMM0003 A Z ISE80 C01L Z ISE80 A2...
Page 59: ...58 PS OMM0003 A Z ISE80 A2L Z ISE80 B2...
Page 60: ...59 PS OMM0003 A Z ISE80 B2L...