8 Electromagnetic compatibility
Operating Instructions and Installation XIOS XG USB module and sensors
8.1 Accessories
båÖäáëÜ
8
Electromagnetic compatibility
XIOS XG complies with the requirements for electromagnetic
compatibility (EMC) according to IEC 60601-1-2.
XIOS XG is hereinafter referred to as "UNIT". Observance of the following
information is necessary to ensure safe operation regarding EMC
aspects.
8.1
Accessories
Designation of the interface cable:
USB cable A/B 2 m with ferrite, REF 64 04 235
The UNIT may only be operated with accessories and spare parts
approved by Sirona. Unapproved accessories and spare parts may lead
to an increased emission or to a reduced immunity to interference.
The UNIT should not be operated in the immediate vicinity of other
devices. If this proves to be unavoidable, the UNIT should be monitored
to check and make sure that it is used properly.
8.2
Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in such
an environment.
System components Manufacturer
Order No.
Serial No.
CDR Remote
SIRONA USA
Not
applicable
10171017
CDR Elite Intraoral
Sensor & Cable
Assembly
SIRONA USA
Not
applicable
20000102
Emission measurement
Conformity
Electromagnetic environment - guidelines
RF emissions according to CISPR 11
Group 1
The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions according to CISPR 11
Class B
The UNIT is intended for use in all facilities,
including residential areas and in any facilities
connected directly to a public power supply
providing electricity to buildings used for residential
purposes.
Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations / flicker according
to IEC 61000-3-3
coincides