5 Operation
Sirona Dental Systems GmbH
5.2 Determine unit status
Operating Instructions and Installation XIOS XG USB module and sensors
5.2
Determine unit status
The XIOS XG USB module features three LED displays. They indicate
the unit status.
Sensor
SIDEXIS
XG /
SIDEXIS 4
Sensor status
LED, green
Exposure
readiness LED,
orange
Sensor
connection LED,
green
Status / bug fixing
Connected Running
at
Flashes every
half to one and
a half seconds
at
Ready for exposure
Connected Running
on
off
on
Exposure in progress
Connected Not running off
off
on
Start SIDEXIS XG / SIDEXIS 4 to
establish unit exposure readiness
Connected Running or
not running
off
on
off
Short circuit or overcurrent condition.
Replace the sensor cable. If the
problem persists, replace the sensor.
Connected Running or
not running
off
on
flashing
Undercurrent condition. Replace the
sensor cable. If the problem persists,
replace the sensor.
Not
connected
Not running off
on
off
Connect the sensor and start
SIDEXIS XG / SIDEXIS 4.
Connected
or not
connected
Running or
not running
On or off
flashing
off
USB voltage supply too low. Use
another original USB cable, another
USB port on the PC or use a hub.