6 Maintenance and inspection
Operating Instructions and Installation XIOS XG USB module and sensors
6.3 Replace the sensor cable
båÖäáëÜ
Unscrew a damaged cable from the sensor
1. Disconnect the sensor from the XIOS XG system.
2. With e.g. a dental instrument remove the protective screw cap from
the rear side of the sensor.
3. Using the screwdriver supplied with the set remove the 2 screws
which fasten the cable to the sensor.
Replace the Elastomer strip
1. Remove the existing Elastomer strip from the sensor with tweezers.
2. Insert a new Elastomer strip in the slot. Check that it is positioned
correctly by pressing lightly with the finger.
The Elastomer strip must be placed exactly in the slot so that the
sensor works.
IMPORTANT
Use the new parts supplied with the cable for the assembly!
IMPORTANT
The spare parts for XIOS XG Select sensors are blue while those of the
XIOS XG Supreme sensors are red. Always replace the existing
Elastomer strip with one of the same color.
The Elastomer strip must be placed exactly in the slot so that the sensor
works.
XIOS XG Supreme
XIOS XG Select