6 Maintenance and inspection
Sirona Dental Systems GmbH
6.3 Replace the sensor cable
Operating Instructions and Installation XIOS XG USB module and sensors
6.3
Replace the sensor cable
The sensor cable can be replaced with a new one in the event of damage.
A "sensor cable" replacement set is required for this, see Consumables
and spare parts"Consumables and spare parts". This contains the
required spare parts and tool:
● Sensor cable
● Screwdriver
● 2 protective screw caps
● 2 Elastomer strips
● 4 flathead screws
NOTICE
The electric plug contacts of the sensor are exposed during installation.
The sensor may be destroyed by the electrostatic discharge at the
contacts.
➢ Before installation discharge at a conductive grounded item, e.g., a
water faucet or a bare heating pipe.
NOTICE
No dirt or moisture must be allowed to get into the exposed plug
contacts.
Dirt causes contacts faults, moisture may lead to a short circuit.
➢ Wash and dry your hands. Do not wear powdered gloves. The
powder could become deposited in the plug contacts of the sensor.
➢ Place the sensor on a clean, dry and stable surface.