5 Operation
Operating Instructions and Installation XIOS XG USB module and sensors
5.1 Enable exposure readiness
båÖäáëÜ
"Exposure readiness " window when using a XIOS XG Supreme sensor
6. When using a XIOS XG Supreme sensor: Select the desired filter to
be applied to the x-ray image produced depending on the indication.
The selection can also be changed after the exposure.
The button for the selected image filter is shown inside a red
frame. The
"Universal" image filter is pre-selected by default.
The orange LED on the USB module flashes, the adjacent LED lights
up green, see "Determine unit status" [ → 36]. The USB module is
now also ready for exposures. The exposure can be released now.
WARNING
Connecting / replacing a sensor on the USB module
You can connect or replace the sensor on the USB module until the
radiation is released. The sensor can also be exchanged when the
exposure readiness is already created.
When you separate the sensor from the USB module, after the radiation
is released, and the image has not yet been transferred, the image data
will be lost.
➢ Do not disconnect the sensor from the USB module before the
image data has been transferred to SIDEXIS XG / SIDEXIS 4.