4 Installation
Sirona Dental Systems GmbH
4.2 Install the PC software
Operating Instructions and Installation XIOS XG USB module and sensors
5. Remove any paint around the drill hole to achieve a good metal
contact.
6. Fasten the M4 connecting screw to the PC casing firmly with a
toothed lock washer and nut.
7. Screw down the protective ground conductor as shown.
8. Affix the "Grounding point" label provided next to the protective
ground conductor connection.
9. Connect the protective ground conductor to a suitable equipotential
bonding conductor.
4.2
Install the PC software
The PC must be in a state of operational readiness before the installation
of the XIOS XG Intraoral System. Make sure that the hardware and the
operating system are properly installed. Refer to the manuals for your PC
and operating system.
The installation of SIDEXIS XG / SIDEXIS 4 is described in the installation
instructions for SIDEXIS XG / SIDEXIS 4.
The SIDEXIS XG / SIDEXIS 4 plug-in for the XIOS XG USB module must
be installed in addition to SIDEXIS XG. To do this, proceed as follows:
✔
SIDEXIS XG / SIDEXIS 4 (SIDEXIS XG 2.5.6 or higher) is installed
on the PC.
1. Log onto your PC with an administrator account.
2. Insert the
"Sirona XIOS XG Select/Supreme 1.0 - Installation" CD
into your PC's CD/DVD drive.
3. The setup normally starts automatically. If the setup does not start,
double-click on the "Autorun.exe" file in the main directory of the
installation CD.
The setup window opens.
NOTICE
The USB module must not be connected to the PC during installation of
the SIDEXIS plug-in for XIOS XG.