5 Operation
Sirona Dental Systems GmbH
5.6 Releasing the exposure
Operating Instructions and Installation XIOS XG USB module and sensors
5.6
Releasing the exposure
✔
The exposure parameters for the X-ray tube assembly are set, see
"Select the exposure parameters for the X-ray tube assembly" [ → 53].
✔
The sensor is positioned in the patient's mouth with the
corresponding auxiliary exposure equipment, see "Position the
sensor" [ → 38].
1. Ensure that the USB module is ready for exposures. The orange LED
on the USB module must flash, the adjacent LED must light up green,
see "Determine unit status" [ → 36].
2. Ensure that SIDEXIS XG / SIDEXIS 4 is ready for exposures. The
green display must be flashing in the Exposure readiness window.
3. Check that the X-ray tube assembly is in the correct position and take
the X-ray exposure.
4. Remove the sensor from the hygienic protective sleeve. For this
follow the instructions in section "Remove the hygienic protective
sleeve from the sensor" [ → 57]. The used sensor holder tab and
hygienic protective sleeve must be disposed of after the examination.
The guide rod and localizer ring must be cleaned and sterilized. The
sensor must be wiped with disinfectant before it is used for another
patient.
5. Following the X-ray exposure, carefully remove the sensor to prevent
it from falling down.
6. Continue with image processing in SIDEXIS XG / SIDEXIS 4.