62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
15
Sirona Dental Systems GmbH
Required materials
Installation Instructions HELIODENT
PLUS
Required tools and auxiliary materials
båÖäáëÜ
● Screwdriver, Torx
®
, sizes T10, T15 and T20
● Slotted screwdriver, sizes 2.5 mm and 3.5 mm (3/32" and 9/64")
● Allen wrench, size 13 mm
● Socket wrench or ratchet wrench (with extension), size 13 mm
Retaining ring pliers, external
● Retaining ring pliers for outside rings (shafts)
Drift punch
● Drift punch
Ammeter
● Multimeter or ammeter (battery-operated)
Bender tester
● Test unit for device leakage current measurement, e.g. Bender tester
or line-frequency, high-resistance measurement voltage source
(isolation transformer) and measuring circuit (MD) that meets the
requirements of IEC 60 601-1.
Power source
● Power source for protective ground wire test
Technical data:
– No-load voltage max. 6V
– Short-circuit current at least 5A - max. 25A