62 14 972 D 3507
10
D 3507
.
031.01.07
.
02 11.2010
Installation options
Sirona Dental Systems GmbH
Installation Instructions HELIODENT
PLUS
empty
Installation option 4
Installation option 4
Release outside of the X-ray room with remote control
● Release
– Manual release S3
or
– Remote control release key S9
empty
Installation option 5
Installation option 5
Release outside of the X-ray room with Remote Timer
● Release
– Manual release S3
or
– Release key on the control membrane S4
empty
Installation option 6
Installation option 6
Release outside of the X-ray room with remote control, door contact sa
-
fety circuit
● Door contact
– Door contact S7 wired to the wall adapter
● Release
– Manual release S3
or
– Remote control release key S9
4
L9
S3
S9
NOTICE
Installation prerequisites
Use of the remote control is permissible only if the yellow X-Ray LED is
visible to the operating personnel during radiation release.
5
L6
S3
S4
NOTICE
Length of cable supplied for Remote Timer approx. 10 meters (394")
(must not be extended).
Conduit int. dia. at least 12 mm (1/2").
6
L9
S7
S3
S9
NOTICE
Installation requirement
Use of the remote control is permissible only if the yellow X-Ray LED is
visible to the operating personnel during radiation release.