Electrical connection
Sirona Dental Systems GmbH
Connecting the release and the door contact
Installation Instructions HELIODENT
PLUS
6.4.4
Installation option 2
empty
Configuration
Installation option 2
Release in the treatment room with remote control
● Release
– Manual release S3
or
– Remote control release key S9
empty
Prerequisites
● The remote control has been installed (see section "Installing the
remote control/remote timer [ ➙ 23]").
● Only with use of the manual release S3 (coiled cable):
– The holder for manual release S3 was attached to the cover of
the remote control. See section "Installation on the protective
cover of the remote control [ ➙ 29]".
empty
Installation of cable L9 (on DX1)
1. Connect cable L9 to connectors X401.1 and X401.4 on PC board
DX1.
2. Secure cable L9 with the shielding clamp (B).
empty
Installation with remote control release button S9
➢ Connect cable L9 and remote control release button S9 to the 3-pin
terminal as shown.
● Cable drawn in black = concealed installation
● Cable drawn in gray = surface installation
2
L9
S3
S9
L9
2
X401.1
X401.2
X401.3
X401.4
S3/S9
DX1
X401
B
L9
NOTICE
The shielding of cable L9 must be in flat contact with the shielding clamp
(B).
L9
2
X401.1
X401.2
X401.3
X401.4
S3/S9
DX1