Starting up
Sirona Dental Systems GmbH
Individual unit adaptation
Installation Instructions HELIODENT
PLUS
7.3
Individual unit adaptation
empty
Explanation
The functionality of the HELIODENT
PLUS
can be adapted to customer
requirements via service routines.
empty
Command
See section "Service routines [ ➙ 68]“.
empty
Overview
The following adaptation of the key assignment and the exposure tables
are most often requested by the customer:
Other service routines enable individual display settings:
More information on the service routines is provided in the Service
Manual
In addition, the user can activate preselected service routines via his own
service area (see operating instructions).
Service routine
Function
S05
Selection of sensor type
S04
Selection of film type
S06
Selection of cone type
S07
Set diaphragm type
S01
Configuration of transparency compensation for
films
S02
Configuration of transparency compensation for
sensors
S13
Activation/deactivation of 60/70 kV toggle
S14
Activation/deactivation of detector medium toggle
(film/sensor)
Service routine
Function
S08
Configuration of radiation time and dose display
after an X-ray exposure
S09
Configuration of time-out time of the radiation
time and dose display
S10
Configuration of display options: Area dose and
actual radiation time
S11
Configuration of power-save mode
S12
Configuration of time-out time of the power-save
mode