User Manual For - CONTROLLER/DATA RECORDER MultiCon CMC-N16
type
range*
resolution
full range
full range accuracy
limited range
limited range
accuracy
[ mV ]
[ μV ]
[
°C
]
[
°C
]
[ % ]
[
°C
]
[
°C
]
[ % ]
K
-10÷120
4
-200÷1370
±8
±0,51
-100÷1370
±4
±0,27
S
-10÷30
2
-50÷1768
±7,5
±0,41
0÷1768
±6
±0,34
J
-10÷120
4
-210÷1200
±6,3
±0,44
-100÷1200
±3
±0,23
T
-10÷30
2
-200÷400
±1,9
±0,31
-100÷400
±1
±0,20
N
-10÷120
4
-200÷1300
±12
±0,80
-100÷1300
±5,9
±0,42
R
-10÷30
2
-50÷1768
±8,6
±0,47
0÷1768
±5,5
±0,31
E
-10÷120
4
-200÷1000
±4,7
±0,40
-100÷1000
±2,7
±0,25
L
-10÷120
4
-200÷800
±4,4
±0,44
-100÷800
±2,5
±0,28
B
-10÷30
2
250÷1820
±15
±0,96
480÷1820
±6
±0,45
C
-10÷120
4
50÷2290
±13
±0,51
50÷1900
±9,2
±0,5
* taking into account the operation of the device in environment temperature range of
-20'C÷50'C.
Tab. 8.3 Thermocouple measurement accuracy for the FUN2 and FUN4 hardware input
configurations in TC mode
type
norm
K
PN-EN 60584-1:2014
S
PN-EN 60584-1:2014
J
PN-EN 60584-1:2014
T
PN-EN 60584-1:2014
N
PN-EN 60584-1:2014
R
PN-EN 60584-1:2014
E
PN-EN 60584-1:2014
L
GOST R 8.595:2004
B
PN-EN 60584-1:2014
C
PN-EN 60584-1:2014
Tab. 8.4 Thermocouple norm for the FUN2 and FUN4 hardware input configurations in TC
mode
259