User Manual For - CONTROLLER/DATA RECORDER MultiCon CMC-N16
type
range
temperature
range
resolution
accuracy
accuracy
[ Ω ]
[ °C ]
[ Ω ]
[ °C ]
[ % ]
Pt100
0÷325
-100÷600
0.01
±1
±0,14
Pt500
0÷3250
-100÷600
0.1
±2
±0,3
Pt1000
0÷3250
-100÷600
0.1
±1
±0,14
Pt'50
0÷325
-200÷600
0.01
±2
±0,25
Pt'100
0÷325
-200÷600
0.01
±1
±0,12
Pt'500
0÷3250
-200÷600
0,1
±2
±0,25
Cu50
0÷325
-50÷200
0.01
±1,8
±0,73
Cu100
0÷325
-50÷200
0.01
±0,9
±0,37
Cu'50
0÷325
-200÷200
0.01
±1,6
±0,4
Cu'100
0÷325
-200÷200
0.01
±0,8
±0,2
Ni100
0÷325
-60÷180
0.01
±0,7
±0,3
Ni500
0÷3250
-60÷180
0.1
±1,4
±0,58
Ni1000
0÷3250
-60÷180
0.1
±0,7
±0,3
Tab. 8.5 RTD measurement accuracy for the FUN2 and FUN4 hardware input configurations
in RTD mode
260