Maintenance, cleaning and transport
TP17 / TPM
- 48 -
© SIKA • Ba_TP17-TPM_en • 01/2016
Silicone oil:
Remove the sensor cage from the tank.
Remove the magnetic stirrer with the help of the magnetic lifter
Clean the cage, the magnetic stirrer and the tank with water to which a generous amount
of rinsing liquid has been added.
Remove the cleaning water to the maximum extent possible using the draining pump.
Dry the tank, the sensor cage, the magnetic stirrer and the draining pump thoroughly.
8.3
Transport
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
Allow the calibrator to cool down before transport.
Carry the calibrator only at the handle.
Observe the following instructions for a safe transport of the TP17 / TPM:
The calibrator has to cool down to room temperature.
The calibrator is switched off and was disconnected from AC power.
All connected sensors and devices have been removed.
The tank of the micro bath calibrator has been closed with the transport cover.
The calibrator must be transported upright.
Avoid bumps and shocks.
Let the calibrator neither fall nor tip over.
Protect the calibrator from inadequate environmental influences.
Use our transport case which is available as accessories.
Carry the calibrator only at the handle.