Maintenance, cleaning and transport
TP17 / TPM
- 46 -
© SIKA • Ba_TP17-TPM_en • 01/2016
Thermal fuse:
The thermal fuses of the calibrator are located on the front side and are integrated in the
mains connection. If there is a mains voltage present, but the display is dark and the fan is
not running, you should check the fuses and replace them if required.
Pull the mains connection cable from the calibrator.
Prise open the fuse compartment from the bottom with a fingernail or a flat screwdriver.
Remove the compartment with the fuses.
Check the fuses and replace the faulty fuses.
IMPORTANT! Only use fuses of the same type (
§ 10.2…10.4 ).
Always replace both fuses, even if only one is defective.
Fit the fuse compartment back in place and connect the mains connecting cable (
§ 4.2
"Electrical connection").
Should the fuses blow repeatedly, there is probably a fault in the calibrator.
In this case, send the calibrator to SIKA for repairs (
§ 7+§ 7.1).
8.1.1
Recalibration
The calibrator is adjusted and tested with measuring equipment in accordance with
recognized national standards prior to delivery.
The calibrator should, depending on the application situation, be inspected at appropriate
intervals on the basis of DIN ISO 10012. We recommend you to return the calibrator to SIKA
at intervals of max. 12 months or approx. 500 operating hours for recalibration and
readjustment.
Recalibration is based on the directive DAkkS-DKD R5-4 of the German Accreditation Body.
The measures described here are applied and considered during recalibration.