TP17 / TPM
Problems
Technical changes reserved
- 43 -
7
Problems
CAUTION! Material damage!
The TP17 / TPM cannot be repaired by the user! In case of a defect, the device must be
returned to the manufacturer for repair.
Never open the TP17 / TPM and perform any repair yourself.
The following table details what problems you can solve yourself and how to solve them.
If you are unable to remedy any particular problem, then immediately disconnect the
calibrator in order to protect it from unintended operation.
Contact your supplier or directly to SIKA. Please send the device for repair with a brief
description of the problem, the environmental conditions and the length of time the device
was operational before the problem occurred.
Problem
Possible cause
Remedy
- - - -
Interruption of the internal
reference sensor or the internal
reference sensor is defective.
The controller switches off the
power supply to the heating
cartridge (servicing required).
uuuu
Measured temperature under the
limit value of the internal
reference sensor (under range -
200 °C)
oooo
Measured temperature above the
limit value of the internal
reference sensor
(over range +850 °C)
ErEP
Possible fault in the EEPROM
memory of the controller
Press the P key
Fan not running
The fan is defective or blocked
The temperature switch is
possibly triggered, switching off
the power supply to the heating
cartridge (servicing required)
End temperature is not
achieved
Solid state relay is defective or the
heating / cooling element has
short circuited or aged
Servicing required
Sensor break.
External reference sensor not
properly connected.
Recheck connection and connect
properly.
Cable break or short circuit.
Service call required.
No display
Controller defective
Servicing required
No function
Network connection not
established correctly or fuse
defective
Check the network connection and
fuse