TP17 / TPM
Commissioning and operation
Technical changes reserved
- 23 -
Tank:
The maximum fill level in the tank is displayed by the upper edge of the aluminium
lining.
The maximum fill level is ~0.45 litres.
Tub insert:
The maximum fill level with the tub insert is below the fixture for the sleeve exchange tool.
The maximum fill level is ~0.32 litres.
4.3.4.3
Filling the micro bath *
1
IMPORTANT! Pay attention to § 4.3.4.2 "Information on filling amounts"!
When filling, leave enough room for thermal expansion, displacement by sensors and
level rise due to stirring.
Screw on the cover of the tank/ tub insert.
Insert the tub insert in the tank using the sleeve exchange tool
(only for tub insert).
Place the magnetic stirrer in the tank / tub insert.
Insert the sensor cage.
Insert the test specimens in the sensor cage.
This accounts for the volume of the sensors to be
tested.
Fill the calibration liquid in the tank / tub insert.
Leave sufficient reserve space for an additional rise in
the level.
If necessary, remove the test specimens again.
Screw the work cover onto the tank and insert the sensors through the work cover into
the tank or tub insert.
*
1
For tubs already filled, some steps are not required.