TP17 / TPM
Operation
Technical changes reserved
- 37 -
ACTIVATE TEMPERATURE PROFILE!
If you have carried out settings for these three settings, the calibrator / micro bath uses
the new values only when modifying the set temperature or switching the calibrator off
and on again.
A further procedure is to switch off the automatic control prior to modifying parameters
and to switch it on again afterwards
(
§ 5.4.1).
The heating and cooling gradients and the duration time can be set in the parameter level
′
rEG
.
Pressing the
P
key for approx. 5 sec., the main menu opens.
The last selected function appears on the upper display.
LED SET flashes on the lower display.
Press the
or
key
until OPEr appears.
Press the
P
key again, the
group level
opens.
OPEr
appears on the upper display.
′
SP
appears on the lower display and LED SET
flashes.
Use the
key to select the group
′
rEG
.
OPEr
appears on the upper display.
′
rEG
appears on the lower display and LED SET
flashes.
Press the
P
key again, the
parameter level
opens.
′
rEG
appears on the upper display.
SLor
flashes on the lower display
Group
′
SP
Group
′
rEG
Parameters for heating gradient SLor