TP17 / TPM
Operation
Technical changes reserved
- 35 -
If no key is pressed for approx. 15 seconds, the device automatically returns to a previous
level up to the
calibration mode
.
Retrieving the saved set temperatures (S version):
The saved set temperatures can be retrieved in calibration mode.
Press the
U
key for approx. 2 sec., the current set
point memory opens.
The current reference temperature appears on the
upper display.
The set point memory SP… appears on the lower
display for 2 sec. followed by the current set
temperature.
To receive another saved set point SP1, SP2, SP3
or SP4, press the
U
key again.
The selected temperature value is immediately
adopted and targeted.
5.4.4
Gradient control / temperature profile (S version)
It is possible to carry out a gradient control yourself and thus determine the time in which the
set temperature is reached. The time can be shorter or longer than the time usually required
by the calibrator / micro bath.
When modifying the set temperature or switching on the calibrator / micro bath it is
automatically determined which of the gradients (heating gradient “SLor” or cooling gradient
“SLoF”) is to be used.
Additionally, you can ensure that the calibrator / micro bath switches to the set temperature
in set point memory SP2 as soon as the set temperature in set point memory SP1 has been
achieved and after a programmed duration time “dur.t”; this creates a simple temperature
profile.
After switching on the calibrator / micro bath the temperature profile is automatically carried
out.
Gradient control and temperature profile
Retrieving the set temperatures display