TP17 / TPM
Technical data
Technical changes reserved
- 53 -
10.4
Characteristics series TP M … S
Series
Characteristics
TP M 165 S
TP M 225 S
TP M 255 S
Calibrator
Temperature range:
-
Dry block
-35…165 °C
T
R
…225 °C
T
R
…255 °C
-
Infrared
-35…165 °C
T
R
…225 °C
T
R
…255 °C
-
Surface
-25…150 °C
T
R
…200 °C
T
R
…200 °C
-
Micro bath(water)
0…100 °C
T
R
…100 °C
T
R
…100 °C
-
Micro bath (silicone oil)
-35…165 °C
T
R
…225 °C
T
R
…255 °C
Accuracy:
-
Dry block
±
0.3 °C
±
0.4 °C
±
0.4 °C
-
Infrared
±
0.5 °C
±
0.5 °C
±
0.5 °C
-
Surface
±
1.0 °C
±
1.0 °C
±
1.0 °C
-
Micro bath
±
0.1 °C
±
0.2 °
±
0.2 °
Emission degree
-
Infrared
0.9994
Stability:
-
Dry block
±
0.05 °C
-
Infrared
±
0.05 °C
-
Surface
±
0.2 °C
-
Micro bath
±
0.05 °C
Display range
-50…165 °C
0…225 °C
0…255 °C
Electrical characteristics
Supply voltage:
-
100…240 V
AC
50/60 Hz
50/60 Hz
50/60 Hz
-
230/240 V
AC
-/-
50/60 Hz
50/60 Hz
-
100…115 V
AC
(optional)
-/-
50/60 Hz
50/60 Hz
Power consumption
375 W
1000 W
1000 W
Fuse:
-
100…240 V
AC
6.3 A slow
10.0 A slow
10.0 A slow
-
230/240 V
AC
-/-
6.3 A slow
6.3 A slow
-
100…115 V
AC
(optional)
-/-
10.0 A slow
10.0 A slow
Serial interface:
-
Serial interface type t
RS 485 optoisolated
-
Communication protocol
MODBUS RTU (JBUS)
-
Serial transfer speed
9600 Baud
Process variables
Test specimen holder:
-
Bore / Depth
∅
60 mm / 170 mm
-
Measurement zone [mm]
123…163 • Infrared: 110 • Surface • Micro bath:110…150
Housing dimensions:
-
Width
-
Height
-
Depth
210 mm
380+50 mm
300 mm
150 mm
330+70 mm
270 mm
150 mm
330+70 mm
270 mm
Weight
~12.5 kg
~7.5 kg
~7.5 kg
Sensor basket working depth
150 mm