Instruction Manual ColorPlus Ex
10210E/5
35
Action
Display (example)
Remarks
1.
Switch to the service
mode.
* 100% ADJUST. *
* *
Section 4.3
2.
* MEAS.RANGE *
* LIMITS *
3.
> LI1 source <
Channel 1
LI1 = Limit 1
LI2 = Limit 2
4.
Activate editing mode
with
.
LI1 source
> Channel 1 <
5.
Select source:
/
LI1 source
> ... <
Channel to which the
limit should apply
(channel 1 .. 3 or math
channel 1 or 2)
6.
Confirm selection with
.
> LI1 source <
...
7.
> Upper limit1 <
1.000 E/m
8.
Activate editing mode
.
Upper limit1
> 1.000 E/m <
9.
Set upper threshold
value:
/
Change number
/
Change place
Upper limit1
> ... <
Set the value at which
the limit should be
activated.
10.
Confirm setting with
.
> Upper limit1 <
...
11.
> Lower limit1 <
0.800 E/m
12.
Activate editing mode
with
.
Lower limit1
> 0.800 E/m <
13.
Set lower threshold
value:
/
Change number
/
Change place
Lower limit1
> ... <
Now set the value at
which the limit should
be deactivated again.
14.
Confirm setting with
.
> Lower limit1 <
...
15.
> LI2 source <
Channel 2
Repeat procedure for
limit 2
steps 6 to 16.
16.
+
(together)
K1 68.2 E/m
K2 85.8 %T
Instrument in normal
operation.