Instruction Manual ColorPlus Ex
44
10210E/5
5.6
Replacement of the desiccant in the receiver
Action
Remarks
1.
Do not open the photometer before the
power supply to the control unit has been
disconnected and then a waiting time of ten
minutes has passed to allow the light source
to cool down sufficiently (EXPLOSION
HAZARD)! (
Cooling down of instrument to a
safe temperature level and fading of the
residual charges)
Disconnect the power supply of the
photometer to the control unit.
2.
Loosen the Allen screw (arrow) at the
receiver enclosure by about ½ turn and pull
off the locking piece.
3.
Loosen nut (A) enough to be able to swing
the hinged ring away easily. Take off the
entire clamping ring and then remove the
receiver enclosure.
4.
Remove the saturated desiccant bag (A).
Take the new desiccant bag, shake its
contents to one end and roll the bag up. In
this form it can be inserted easily into the
receiver.
5.
Check the enclosure seal (B) and replace it if
necessary.
6.
Reassemble the photometer in reverse order
and restart it.