Instruction Manual ColorPlus Ex
10210E/5
15
3.2
Photometer installation in inline housing
The photometer can be installed using a standardized inline housing (Varivent®
or equivalent) in either vertical or horizontal product pipes.
The groove (A) on the transmitter and receiver must be installed pointing
upward!
Be sure to install the ColorPlus Ex so the transmitter and receiver are
positioned horizontally opposite each other! Only install a housing that
makes this possible!
The photometer must be installed in the pipe at least 2 m away from sight
glasses or other sources of stray light.
Figure 7: Installation in vertical product
pipes.
Figure 8: Installation in horizontal product
pipes.
Installation position