Instruction Manual ColorPlus Ex
10210E/5
11
Operating voltage:
24 V DC (from control unit)
Interfaces:
Modbus / with optional bus coupler Profibus DP
Flow cell
-
bypass flow cell
Stainless steel 1.4435,
connections G ¼” female thread
-
inline flow cell
Varivent® housing of stainless steel 1.4404,
DN 40 – 150.
-
user-specific
(inline flow cell)
Titanium, Hastelloy, PVDF etc.
Enclosure:
Aluminum AlSi1MgMn, painted
Dimensions:
Detailed dimension drawing: see Section 11
Weight:
About 10 kg (not including flow cell)
Degree of protection:
IP65
Type of Ex
protection/temperature
class:
Ex d IIC T3/T4/T5/T6 Ga/Gb, temperature class
dependent upon the temperature of the medium:
Medium temperature
Temperature
class
-20 .. +80° C
T6
-20.. +95° C
T5
-20.. +130° C
T4
-20 ..+195° C
T3
Operating voltage:
85 .. 264 V; 47 .. 63 Hz or 24 VDC; 25 W
Space required:
SIREL:
200 mm x 157 mm x 96 mm
SIREL robust: 220 mm x 160 mm x 90 mm
Detailed dimension drawing: see Section 11
Weight:
SIREL:
about 1.5 kg
SIREL robust: about 2.0 kg
Degree of protection:
IP65
Connections:
0/4 .. 20 mA, max. 600
, max. 24 V, with
electrical isolation, max. 50V against earth
Relay contacts max. 250 VAC, max. 4 A
Digital inputs and outputs max. 5 V
Photometer