Instruction Manual ColorPlus Ex
10210E/5
53
The ColorPlus Ex can detect the following faults:
Message
Signifies
Possible causes
connection
lost ...
The control unit has no
connection to the photometer.
Connection to
photometer interrupted
Defective electronics
(
Service technician)
Fault
UV-lamp
The detectors are not receiving
any light.
Defective UV light
source (
Section 5.9)
Defective electronics
(
Service technician)
Fault
LED
The detectors are not receiving
any light from the LED.
Defective LEDs
(
Service technician)
Fault
Measurement
The AD converter is overloaded.
The detectors are
receiving too much light
Enclosure not tight
Defective electronics
(
Service technician)
Fault-
Current 1
Incorrect reading being
measured at reading output 1.
Open connection
terminal at reading
output (
Section 3.7)
Interruption in control
loop of reading output
Loose contract at
reading output
Fault-
Current 2
Incorrect reading being
measured at reading output 2.
Open connection
terminal at reading
output (
Section 3.7)
Interruption in control
loop of reading output
Loose contact at
reading output
Fault
AnalogIn 1
Analog signal 1 has fallen below
a preset threshold
Signal below the
configurable fault
threshold
Fault
AnalogIn 2
Analog signal 2 has fallen below
a preset threshold
Signal below the
configurable fault
threshold
System fault
At the program start, at least 5
system errors were entered in
the history within two minutes
(Reference Handbook)
The parameter memory
has lost its data (
Service technician)
The error history is
overfilled (
Service
technician)
Table 6: Fault messages