Instruction Manual ColorPlus Ex
26
10210E/5
3.8
Initial start-up
Carry out the initial start-up according to the following table. In the event of
malfunctions, consult Section 6.
Action
Remarks
1.
Make sure the photometer and
control unit are correctly mounted
and connected.
Section 3
2.
Switch on the power supply to the
control unit.
After a few seconds, a reading will
appear on the control unit display
(depends on the instrument type and
which unit is set).
3.
Set the language of your region
(
Section 4.4).
Now the menu text appear in the
desired language.
4.
Set the measuring range you
require for your measurement
conditions (
Section 4.5).
5.
Set the relay functions you require
for your measuring duty
(
Section 4.7).
6.
Secure your settings against
tampering with an access code
(
Section 4.9).
If you don’t require an access code,
skip this step.
7.
Carry out a 100% adjustment.
Section 5.2